16
3. Mixing (Pipetting, Diluting)
Mixing ON = Mixing is currently OFF and pressing Mixing ON will turn Mixing 
ON
Mixing OFF = Mixing is currently turned on and pressing Mixing OFF will turn 
Mixing OFF
When mixing is activated, pipette will show the default volume for mixing 
(80% of the dispensed volume). Mixing volume can be adjusted by the user by 
turning the adjustment wheel and the settings can be confirmd by pressing 
the operating button, mixing volume cannot exceed the pipette’s maximum 
volume.
To change the mixing from manual to automated, press the operating button.
Manual mixing: hold the operating button pressed down.
Set the mixing times by turning the adjustment wheel and press the 
operating button to confirm the setting. Automated mixing can be paused by 
pressing the operating button. Press the operating button again to continue 
dispensing. Press QUIT (left soft key) to stop mixing and empty the tip by 
pressing the operating button.
4. Excess Adjustment (Rev. Pipetting, Multi Dispensing, Sequential 
Dispensing)
Excess Adjustment can be used to set the excess volume. When activated, the 
excess volume is set to default. Turn the adjustment wheel to reset the excess 
volume and confirm the setting by pressing the operating button.
NOTE! Default is the maximum excess volume and can only be decreased 
from default.
5. Automated Dispensing (Multi Dispensing)
Automated dispensing dispenses the liquid automatically without the need 
to press the operating button. The delay for the dispensing can be set in time 
frame from 0.1 sec. to 9.9 sec.
6. Fast dispensing (Titrate)
Fast volume dispensing dispenses the first volume rapidly. After dispensing the 
fast volume, dispensing is continued manually.
3.3.10 Setup
Setup is the menu for user settings for calibration adjustment, sound and 
backlight, user identification, GLP info and software reset function.
 
3.3.10.1 Calibration Adjustment 
Adjustment enables the user to adjust the pipette in 1–3 calibration points.
1-point adjustment: calibration in one point, set the calibration volume
2-point adjustment: calibration in 10% and 100% of the nominal volume
3-point adjustment: calibration in 10%, 50% and 100% of the nominal 
volume
Adjustment is always done in P mode and the adjustment is valid in all modes 
when activated. When the Adjustment is activated ADJ – symbol is seen on 
display.
–  Select Menu
–  Select Setup
–  Select Adjustment