7. Troubleshooting
7.1 Hardware reset
If the pipette is unresponsive it is possible to force a turning off and reset
reset the hardware. This function will not affect the saved settings (e.g adjust-
ment, memory). To turnoff an unresponsive pipette press simultaneously ON/
OFF button and soft key for few seconds. Pipette will turn off and can then be
turned on again by pressing ON/OFF button.
NOTE! Make sure the pipette is not placed on the charging unit or connected
to the USB cable while being reset.
7.2 Troubleshooting guide
Problem Possible Cause Solution
Leakage
Incomaptible tip
Tip is loose
Pipette is dirty
Pipette is broken
Use original Sartorius Biohit tips
Attach the tip firmly
Clean the pipette
Replace the broken parts or send the pipette for service
Inaccuracy Pipette is dirty
Pipette is broken
Clean the pipette
Replace the broken parts or send the pipette for service
Pipette does not turn on Pipette is in OFF mode
Empty battery
Turn the pipette ON by pressing ON/OFF button
Connect the pipette to the charger
Piston is jammed Inner lower parts are loose
Pipette is dirty
Pipette is broken
Open the lower part and tighten the parts
Clean the pipette
Replace the broken parts or send the pipette for service
Pipette does not aspirate
liquid properly
Safe-Cone Filter is dirty
Pipette is dirty
Pipette is broken
Replace the Safe-Cone Filter
Clean the pipette
Replace the broken parts or send the pipette for service