11
HINT! Hold the START button down to reaspirate the liquid without emptying 
the tip! Or press NO (left soft key) to continue pipetting without emptying the 
tip.
Additional features
Counter, Excess adjustment and Tracker can be used in conjunction with 
Reverse Pipetting.
–  Press ADV (the right soft key) to activate advanced functions.
–  Turn the adjustment wheel to move to the desired function and confirm the 
selection by pressing the operating button or OK.
–   Counter  counts the number of dispensings. Counting can be selected to 
start in any wanted number.
–  Excess adjustment can be used to set the excess volume. When activated 
the excess volume is set to default. Turn the adjustment wheel to reset 
the excess volume and confirm the setting by pressing the operating 
button. 
Default is the maximum excess volume and can only be decreased from 
default.
–  Tracker advises the user in microplate dispensing by displaying the 
location for the next dispensing.
NOTE! Only one additional feature can be selected at a time.
3.3.3  Multi Dispensing 
Aspirates the total volume as well as an excess volume then repetitively 
dispenses equal volumes of liquid. Recommended in long pipetting series and 
microplate dispensing.
–  Select Multi Disp. in menu. The Multi Disp. mode is activated with the latest 
settings.
–  Start pipetting by pressing the operating button.
– Editing the mode:
–  Press EDIT button (the middle soft key) or turn the adjustment wheel to 
the end. Parameter background colour is highlighted to indicate that 
the editing mode is activated.
–  Turn the adjustment wheel to set the volume and aliquot.
–  Press NEXT (middle soft key) to move the cursor to the next parameter 
to be edited.
–  Confirm the setting by pressing the operating button or OK.
–  Press BACK to return to the previous screen without saving the changes.
HINT! Hold the operating button down to reaspirate the liquid without 
emptying the tip! Or press NO in left soft key to continue pipetting without 
emptying the tip.
Additional features
Timed Automated Dispensing, Excess adjustment and Tracker can be used in 
conjunction with Multi Dispensing.
–  Press ADV (the right soft key) to activate extra features.
–  Turn the adjustment wheel to move to the desired feature and confirm the 
selection by pressing the operating button or OK. 
–  Excess adjustment can be used to set the excess volume. When activated 
the excess volume is set to default.
–  Turn the adjustment wheel to reset the excess volume and confirm the 
setting by pressing the operating button. Maximum excess volume is set 
as default and can only be decreased from default. 
–  Automated dispensing dispenses the liquid automatically without the