24
NOTE! Systematic error is the difference between the dispenced volume
and the selected test volume. Random error is the scatter of the dispensed
volumes around the mean of the dispensed volume (ISO 8655-1)
NOTE! Sartorius Biohit specifications are achieved in strictly controlled
conditions (ISO 8655-1). Users should establish acceptable maximum
permissible errors based on the field of use and the accuracy requirements
placed on the pipette (ISO 8655-1).
4.4 Replacing the battery
It is recommended that the Picus battery is replaced only by authorised
Sartorius service provider. Please contact your nearest Sartorius
representative.
4.5 Replacement parts
To replace broken or worn parts Sartorius Biohit offers a full range of spare
parts.
Always contact the manufacturer or manufacturer’s authorised representative
for original spare parts for your Sartorius Biohit pipette.
4.6 Storage
When not in use it is recommended that the Biohit Picus electronic pipettes
are stored in the charging unit. During long storage periods (several months)
it is recommended that the charging unit is disconnected from the power
outlet and the pipette is turned off by pressing the power button on the top
of the pipette.
Z-values (µl/mg)
Temp. (°C) Air Pressure (kPa)
95 100 101.3 105
20.0
20.5
21.0
21.5
22.0
22.5
23.0
23.5
1.0028
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0036
1.0028
1.0029
1.0031
1.0032
1.0033
1.0034
1.0035
1.0036
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0035
1.0036
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0036
1.0037