Model BP 410 S BP 310 S BP 310 P BP 150
Weighing range structure SuperRange SuperRange PolyRange
Weighing capacity g 410 310 60/120/310 150
Readability g 0.001 0.001 0.001/0.002/ 0.001
0.005
Tare range (by subtraction) g –410 –310 –310 –150
Reproducibility (standard deviation)* g ≤±0.001 ≤±0.001 ≤±0.001/ ≤±0.001
0.001/
0.003
Linearity g ≤±0.002 ≤±0.002 ≤±0.002/ ≤±0.001
0.002/
0.005
Response time (average) s ≤1.5
Allowable ambient operating temperature °C 0 ... +40
Operating temperature range °C +10 ... +30
Sensitivity drift within +10 °C ... +30 °C /°C ≤±2·10
–6
Pan size mm Ø 115
Dimensions (WxDxH) mm 204 x130 x297
Net weight, approx. kg 3.1
External calibration weight
(of at least accuracy class...) g 200 (F1) 200 (F1) 100 (F1) 100 (F1)
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency Hz 48– 63
Power consumption (average) VA 16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx. h 25
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
Round glass draft shield x x x
AC adapter, varies acc. to country x x x
Level indicator x x x
Hanger for below-balance weighing x x x
* = standard deviation of the reproducibility acc. to DIN 8120, Part 3
70