Model BP 8100 BP 6100 BP 4100 BP 2100
Weighing capacity g 8,100 6,100 4,100 2,100
Readability g 0.1 0.1 0.1 0.1
Tare range (by subtraction) g 8,100 6,100 4,100 2,100
Reproducibility (standard deviation)* g ±0.05 ±0.1 ±0.1 ±0.1
Linearity g ±0.1 ±0.1 ±0.1 ±0.1
Response time (average) s 1
Allowable ambient operating temperature °C 0... +40
Operating temperature range °C +10... +30
Sensitivity drift within +10 °C ... +30 °C /°C ±4·10
–6
±5·10
–6
±5·10
–6
±5·10
–6
Pan size mm 210 x 180
Dimensions (WxDxH) mm 238 x297x81
Net weight, approx. kg 3.5 3.0 3.0 3.0
External calibration weight
(of at least accuracy class...) g 5,000 (F2) 5,000 (F2) 2,000 (F2) 1,000 (F2)
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency Hz 48 63
Power consumption (average) VA 16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx. h 25 45 45 45
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
AC adapter, varies acc. to country xxxx
Level indicator xxxx
Hanger for below-balance weighing xxxx
* = standard deviation of the reproducibility acc. to DIN 8120, Part 3
74