Model BP 16000 S BP 12000 S BP 34000 P BP 34 BP 8
Weighing range structure SuperRange SuperRange PolyRange
Weighing capacity kg 16 12 8/16/34 34 8.1
Readability g 0.1 0.1 0.1/0.2/0.5 1.0 1.0
Tare range (by subtraction) kg 16 12 34 34 –8.1
Reproducibility g ±0.1 ±0.1 ±0.1/0.2/ ±0.5 ±0.5
(standard deviation)* ±0.5
Linearity g ±0.2 ±0.2 ±0.3/0.3/ ±0.5 ±1
0.3
Response time (average) s 1.5 1.5 1.5 1.5 1
Allowable ambient
operating temperature °C 0 +40
Operating temperature range °C +10+30
Sensitivity drift
within +10 °C ... +30 °C /°C ±2·10
–6
±2·10
–6
±2·10
–6
±2·10
–6
±10 ·10
–6
Pan size mm 417 x 307 417 x 307 417 x 307 417 x 307 210 x180
Dimensions (WxDxH) mm 307x538x121
Net weight, approx. kg 15 15 15 15 3,0
External calibration weight
(of at least accuracy class...) g 10,000 (F1) 10,000 (F1) 10,000 (F2) 10,000 (F2) 5000 (F2)
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC, –20% ... +15%
(IP 20 protection)
Frequency Hz 4863
Power consumption (average) VA 8 maximum; 10 average
Hours of operation with fully
charged YRB 06 Z
external battery pack, approx. h 28 28 28 28 25
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
AC adapter, varies acc. to country x x x x x
Level indicator x x x x x
* = standard deviation of the reproducibility acc. to DIN 8120, Part 3