Specifications for Basic
plus
Verified Balances with an EC Type Approval
Model BP 211D-0CE BP 301S-0CE
Type BC BC 100 BC BC 100
Accuracy class* kk
Weighing range structure DualRange SuperRange
Maximum capacity Max* g 80/210 303
Scale interval d* mg 0.01/0.1 0.1
Verification scale inter val e* g 0.001 0.001
Minimum capacity Min* g 0.001 0.01
Tare range (by subtraction) g –210 –303
Range of use according to
Directive No. 90/384/EEC* g 0.001–210 0.01–303
Response time (average) s ≤12/3 ≤ 3
Operating temperature range °C +15...+25
Pan size mm Ø 80
Weighing chamber height (effect. dimens.) mm 225
Net weight, approx. kg 6.6 5.5
Selectable weight units g, mg
External calibration weight
(of at least accuracy class…) g 200 (E2) 200+100 (E2)
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency Hz 48–63
Power consumption (average) VA 16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx. h 25 32
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, odd;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
Dust cover x x
AC adapter; varies acc. to country x x
Built-in, motorized calibration weight x x
Level indicator x x
* = EC Directive No. 90/384/EEC for Non-Automatic Weighing Instruments applicable in the
EC/EU Member States
76