Model BP 150-0CE BP 110-0CE BP 610-0CE
Type BA BC 200, BA BC 200,
BD BC 200 BD BC 200 BD BC 200
Accuracy class KKK
Weighing range structure SuperRange SuperRange PolyRange
Maximum capacity Max* g 150 110 610
Scale interval d* g 0.001 0.001 0.01
Verification scale interval e* g 0.01 0.01 0.1
Minimum capacity Min* g 0.02 0.02 0.5
Tare range (by subtraction) g –150 –110 –610
Range of use according to
Directive No. 90/384/EEC* g 0.02–410 0.02–110 0.5 – 610
Response time (average) s ≤1.5
Allowable ambient operating temperature °C 0 ...+40
Operating temperature range °C +10...+30
Pan size mm Ø 115
Dimensions (WxDxH) mm 204 x297x130
Net weight, approx. kg 3.1
Selectable weight units g, kg
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency Hz 48– 63
Power consumption (average) VA 16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx. h 25
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, odd;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
Round glass draft shield x x x
AC adapter; varies acc. to country x x x
Built-in, motorized calibration weight x x x
Level indicator x x x
* = EC Directive No. 90/384/EEC for Non-Automatic Weighing Instruments applicable in the
EC/EU Member States
79