EasyManua.ls Logo

Sartorius BP 3100 P-0CE - Page 78

Sartorius BP 3100 P-0CE
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model BP 410 S-0CE BP 310 S-0CE BP 310 P-0CE
Type BA BC 200, BA BC 200,
BD BC 200 BD BC 200 BD BC 200
Accuracy class KKK
Weighing range structure SuperRange SuperRange PolyRange
Maximum capacity Max* g 410 310 60/120/310
Scale interval d* g 0.001 0.001 0.001/
0.002/0.005
Verification scale inter val e* g 0.01 0.01 0.01
Minimum capacity Min* g 0.02 0.02 0.02
Tare range (by subtraction) g –410 –310 –310
Range of use according to
Directive No. 90/384/EEC* g 0.02–410 0.02–310 0.02–310
Response time (average) s 1.5
Allowable ambient operating temperature °C 0 ...+40
Operating temperature range °C +10...+30
Pan size mm Ø 115
Dimensions (WxDxH) mm 204 x297x130
Net weight, approx. kg 3.1
Selectable weight units g, kg
Power requirements V~ Via AC adapter: STNG 6, 230 VAC or 115 VAC,
–20% ... +15% (IP 20 protection)
Frequency Hz 48 63
Power consumption (average) VA 16 maximum; 8 average
Hours of operation with fully charged
YRB 05 Z external battery pack, approx. h 25
Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, odd;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
Round glass draft shield x x x
AC adapter; varies acc. to country x x x
Built-in, motorized calibration weight x x x
Level indicator x x x
* = EC Directive No. 90/384/EEC for Non-Automatic Weighing Instruments applicable in the