Operating Instructions EVO1S1 | SPO1S1 25
Operation
Level 1 Level 2 Level 3 Level 4
SETUP
t Press !
| - - -
APP.PRG.
t Press O / P, select APP.PRG.
| - - -
UNIT 2
t Press !, press O / P, select UNIT 2
| |- - - o
Grams
t Press !, press O / P, select setting (e. g. GRAMS)
| |- - -
PT.P.LB
t Press !, “o” appears indicating the desired setting is applied.
| - - -
DEC 2
t Press §
|- - -
ALL
t Press O / P, select DEC 2
|- - - o
PolyR.
t Press !, press O / P, select setting
t Press !, “o” appears indicating the desired setting is applied.
t Press § several times to exit the menu
6.7.2.5 Activating the “LOCK” Function
The “LOCK” function protects the scale from unauthorized use. When this function is active, the scale readout shows weight values
only when there is active communication between the scale and a PC. If communication is interrupted, the readout goes blank
and the display shows a padlock symbol. Activation of the LOCK function is configured under “EXTRAS.”
t Accessing the Setup menu (see Chapter 6.7.2.1, page 23).
Level 1 Level 2 Level 3 Level 4
device
t Press O / P, select DEVICE, press !
| - - -
EXTRAS
t Press O / P, select EXTRAS, press !
| - - -
LOCK
t Press O / P, select LOCK, press !
|- - - o
On
t Press O / P, e. g. select ON
|- - -
Off
t Press !, “o” appears,
y The “LOCK “ function
is activated.
t Press § several times to exit the menu
6.7.2.6 Password Request
In addition to the “LOCK” function, you can also configure password protection for additional security.
With this feature, the “LOCK” function can only be deactivated “OFF” by entering the password you configure.
Entering the Password
When the password is requested, the numbers 1 2 3 4 5 6 appear on the display.
You can switch to the numbers 7 8 9 0 and back again via the O / P keys.
t Enter the password by entering the individual numbers on the display.
y When you enter the correct password, then the “LOCK” function will be disabled.