Operating the Scale
Preparation
Configure Parameters for Calibration and Adjustment
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. Select the Setup menu M
3. Select the Balance Functions menu menu soft key w/[TOGGLE]
4. Set parameters for:
Calibration key function 1 9
Calibration/adjustment
sequence 1 10
isoCAL self-calibrating and
adjustment function 1 15
See the chapter entitled “Setup” x y Y X
5. Exit the Setup menu M
info--menu--input