EasyManua.ls Logo

Sartorius FB - Net-Total Formulation Application; Net-Total Purpose, Features, Factory Settings, and Preparation; Net-Total Settings and Additional Functions

Sartorius FB
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Net-Total Formulation R
Purpose
With this application program
you can weigh in different
components up to a defined total.
You can also print out the total
weight and the individual weights
of the components.
Available Features
Taring
Weighing different components
(maximum: 99 components) from
“0” to a defined total component
weight
Storing component weights
(“Store xx comp.”), with
display zeroed after value
stored, and
automatic printout
(print application parameters);
either
of the last component weight
(net value) or
of the total weight (tare value)
Display of the transaction counter
“XXth” (referring in each case
to the next component) in the soft
key label display
Clearing of the component
memory when the weighing
series is canceled c
and printout of the total weight
if you have configured the scale
for GMP-compliant printouts;
otherwise, printout of the net
value p/[PRINT]
Toggling between component
weight and total weight by
pressing w/[TOGGLE]
ISO/GLP-compliant printout
of the total of the individual
component weights (Tot.cp)
Factory Settings
Automatic printout when component
value stored: print net total (7 3 1)
Preparation
Turn on the scale
> All display segments
light up briefly
Select the Net Total application
in the Setup menu: Press M
Select menu:
Press the menu soft key
(w/[TOGGLE])
Set parameter 2 1 6:
See “Configuring the Scale”
Exit the Setup menu:
Press M
Setting Parameters
for the Net-Total Application
k Access the Setup menu:
Select Menu
Set the parameter for automatic
printout when components are
stored
7 3 Print net total
See “Configuring the Scale”
Exit the Setup menu:
Press M
Additional Functions
In addition to functions for:
alphanumeric input,
taring
(not during alphanumeric input),
printing,
you can also access the following
functions from this application:
calibration
(not during alphanumeric input),
setup,
turning off the scale.
Calibration/Adjustment
Press [ISOTEST]/[TEST]
> See “Calibration/Adjustment”
for further instructions
Setup (setting parameters)
Press M
> See “Configuring the Scale” for
further instructions
Turning Off the Scale
Press e/[ON/OFF]
> The scale shuts off
> The display goes blank

Related product manuals