Setting the Drying Parameters
Purpose
Adapt the moisture analyzer’s software to the special requirements of products.
Parameters can be configured individually for each program.
Drying Parameters (Overview)
Setup Input
Manage Load programs
programs*
Change Short name Short name Alphanum.
current Additional identification Long name Alphanum.
program ID 1 Alphanum.
ID 2 Alphanum.
Microwave mode Microwave mode High range
Low range
Analysis mode Analysis mode Standard
Addition of water
Concentration (mg/l)
Drying process Standard Level 2 to 100%
2 stages Level 1** 2 to 100% (30 to 100%)
Time 1 0.0 to 10.0 min
Level 2** 2 to 100% (30 to 100%)
3 stages Level 1** 2 to 100% (30 to 100%)
Time 1 0.0 to 10.0 min
Level 2** 2 to 100% (30 to 100%)
Time 2 0.0 to 10.0 min
Level 3** 2 to 100% (30 to 100%)
Initial weight Off
Minimum/maximum Start Unlocked
Locked
Min 0.01 g to 69.99 g
Max. Min. + 0.01 g to 70.00 g
Target value/tolerance Start Unlocked
Locked
Target value 0.01 g to 63.00 g
Tolerance 1.0 to 50%
Sample capturing Standard
Fast
Very fast
Key without stability
21