  Safety Information
  Note:
  Read these operating instructions carefully before putting  
the LMA200PM into operation. Improper use or handling of the 
analyzer can result in damage and/or injury. The LMA200PM 
may be installed and operated by qualified personnel only.  
Make sure you observe the warnings and safety precautions in 
their entirety during installation and operation, as well as while 
performing maintenance and repair work on the equipment.  
The relevant laws, standards, regulations, guidelines and envi-
ronmental protection provisions applicable in your country must 
also be observed. It is important that all personnel who operate 
the equipment understand these warnings and safety precau-
tions, and have access to the relevant documents at all times. 
Furthermore, the warnings and safety precautions supplied with 
any electrical equipment connected, such as accessories, must be 
observed.  
These warnings and safety precautions may have to be supple-
mented by the equipment operator. Make sure all operating  
personnel are informed of any additions to these instructions.  
Make sure the equipment is accessible at all times.  
If installation or operation is not performed in accordance with 
these instructions and acknowledged technical standards, all 
claims under the manufacturer’s warranty are forfeited.
  Warnings
! The LMA200PM weighs approximately 22 kg (48.4 lbs).  
Two people are required for transport and installation.  
Make sure to comply with the applicable safety regulations.
! Always disconnect the LMA200PM from power before moving it.
! Set up the LMA200PM on a stable, level, even surface.
! Do not operate the LMA200PM outdoors. Do not use this  
equipment in areas subject to safety hazards. If you use electrical 
equipment in installations and under ambient conditions subject 
to stricter safety standards than those described in this manual, 
you must comply with the provisions as specified in the applica-
ble regulations for installation.  
Chemicals, including gases and dusts, that can corrode the inside 
or outside of the analyzer housing or cables must be kept away 
from the equipment.
–  The LMA200PM must not be exposed to strong electromagnetic 
fields (for example, through close proximity to cables carrying 
high electrical loads). Do not install the analyzer in an area in or 
near systems or equipment with magnetic materials.
–  Do not set up the LMA200PM near flammable materials or gas 
lines. 
–  Do not expose the LMA200PM to aggressive chemical vapors or 
to extreme temperatures, moisture, shocks, or vibration. 
! Refer to the chapter entitled “Specifications” for important tech-
nical data. 
–  Make sure the place of installation is adequately ventilated  
to prevent build-up of excessive heat. Leave at least 20 cm 
(about 8 in.) space between the analyzer and the wall and about 
1 m (3 ft.) above the analyzer.  
Make sure the cooling fans are always operational. In the event 
of defect, switch off the moisture analyzer and inform service 
personnel.
Warnings and Safety Precautions
! If you use cables purchased from another manufacturer,  
check the pin assignments in the cable against those specified 
by Sartorius before connecting the cable to Sartorius equipment, 
and disconnect any wires that are assigned differently. The braided 
shield in the cable must be made of metal and must provide at 
least 80% coverage. Both the plug and socket of all connections 
on the cable ends or in the equipment housing must have foil 
shielding that is suitable for high frequencies in accordance with 
CE regulations. The operator shall be solely responsible for any 
damage or injuries that occur when using cables not supplied  
by Sartorius. 
! Disconnect the LMA200PM from power (unplug it from the  
wall outlet) before opening the equipment housing and before 
connecting or disconnecting peripheral devices to or from the 
analyzer.
! A power cord is included in delivery of your LMA200PM. 
  In many European countries, the LMA200PM is electrically 
grounded by the power cord when it is plugged into the electri-
cal socket (mains). If the mains power supply in your area does 
not have a grounding conductor, take the required measures to 
ensure that the analyzer is grounded. Disconnecting the ground 
conductor is prohibited.
! If there is visible damage to the equipment or power cord, turn 
off the power and disconnect the equipment from AC power 
immediately. Lock the equipment in a secure place to ensure that 
it cannot be used for the time being.
! The LMA200PM may be opened only by trained service techni-
cians. 
  Unplug the power cord from the wall socket (mains) before  
performing any work on the equipment.
! Use only original Sartorius spare parts.
  Safety Precautions
! Before switching on the analyzer, measures must be taken to 
prevent excessive exposure to microwave radiation:
  –   Do not wedge any object between the analyzer cover and the 
sample chamber, as the cover provides a protective seal to 
limit electromagnetic emissions
  –   Make sure the sample chamber is completely closed when the 
analyzer cover is lowered
  –   Do not operate the moisture analyzer unless the cover, hinge, 
seal and safety lock are undamaged and fully functional
 
! Do not operate the LMA200PM with the cover open. 
Use of the moisture analyzer for anything other than its intended 
use as described by the manufacturer in this manual may impair 
the proper functioning of the analyzer’s safety features and  
present a hazard for the operator. 
–  The LMA200PM may be operated only by trained and qualified 
personnel. 
! Do not use flammable, explosive, corrosive or polar solvents as 
sample materials. Do not perform moisture analysis on materials 
containing coal, graphite, or metal. Substances that contain  
poisonous or toxic components may only be dried if certain  
precautions are taken; for example, drying under a fume cover. 
The lower toxic limit stipulated for places of work must not  
be exceeded.