Configuring Parameters
in the Setup Menu
Purpose
To adapt the moisture analyzer to meet
individual requirements by selecting
predefined menu parameters in the
Setup menu.
The following menu levels are
password-protected:
Manage programs
Configuration
Calibration
Security management
To open these menus, a PIN must be
entered and the user must have the
required access privileges.
Features
The parameters in the Setup menu are
grouped as follows
(highest menu level):
Modify user data*
Evaluation of results
Configuration*
Calibration*
Security management*
Info
Manage programs*
The parameters in the “Manage pro-
grams” menu are grouped as follows:
Load program
Change current program
Create new program
Delete program
Print current program
Print program list
Print all programs
For details on these options,
please refer to the chapter entitled
“Operation.”
Evaluation of Results
The parameters under “Evaluation of
results” are grouped as follows:
View/print results
Statistics on selected results
Select results
* Access rights required
13
You can save up to 999 results
datasets in the results database.
Cancelled analysis runs are not saved.
Once the database contains 999 data-
sets, a permanent error message
is shown and new datasets cannot be
stored until all existing datasets have
been deleted (see “Factory Settings”
under “Security Management” for
more details).
Each dataset contains the following
fields:
User name
Program name (short form)
Sample ID
Initial weight
Result as ending/starting weight
(i.e., without weight unit)
Display mode
Analysis time
Date and time
View/Print Results
Results are displayed as follows:
Each dataset is shown on two lines:
First line: Internal number, program
name, date, time
Second line: User name, sample ID,
result with unit, and analysis time
There are a number of options for
printing results:
Print all datasets (Prt. all)
Print only the datasets selected for
statistical evaluation (Prt. ST; see
“Select Results” in the next column).
Statistics on Selected Results
The following statistical values are
displayed:
Program (short name)
Number of samples
Mean
Standard deviation
Relative standard deviation
Minimum value
Maximum value
Printing Statistics
Press Print
Printout can be configured
Important note:
Statistics related to a program short
name are calculated irrespectively of
interim program changes;
Statistics are generally derived from
uncalculated results and preferably dis-
played as a percentage of solids (%S*),
which can be toggled to a percentage of
moisture (%M*);
Additional readout: Only not converted
results or Mixed results;
Exception:
Statistics derived from converted results
will be shown in the calculated unit
%M* or %S* if all results are shown in
this unit;
Additional readout: Converted results:
%M* or Converted results: %S*
Select Results
You can choose from a number of crite-
ria in selecting datasets for evaluation.
Selected datasets are highlighted in
green.
The following options are available:
Specify a program name (short form)
Specify the program name and a time
period within which the desired datasets
were stored
Specify the program name and the
name of the user who generated the
desired datasets
Specify the program name and then
select the desired datasets from the
resulting list
You can combine the criteria outlined
above, with the exception of the
individual selection from a list.
See the next page for examples.