Blocking the Numeric Keys
You can block the numeric keys by setting the appropriate menu code.
Numeric keys Code
Blocked 2 5 1*
Accessible 2 5 2
Important Note:
A control command can be input via the balance interface to block the function
and numeric keys (except for e). For more information, see part 4.
Power-On Mode
Depending on the operating mode – line current, battery operation
or continuous operation – you can change the power-on mode of your balance.
The factory setting is: (power) off —> on <—> standby.
In the setting “toggle between on and standby,” the balance power will
turn back on automatically after a power failure has occurred, or after you have
disconnected your balance temporarily from line current.
In the setting “automatic power-on,” the balance will turn back on automatically
after a power failure has occurred or after the balance has been disconnected,
then reconnected to line current. In this setting, the balance can no longer
be turned off by the e key.
Power-on mode Code
(Power) off
> on <
> standby 8 5 1*
On <
> standby 8 5 3
Automatic power-on 8 5 4
Display Backlighting
Display backlighting Code
On 8 6 1*
Off 8 6 2
* = factory setting
2–22