Do you have a question about the Sartorius Micro MC Series and is the answer not in the manual?
Linearity | ±0.002 mg to ±0.02 mg |
---|---|
Interface | RS232, USB, Ethernet |
Calibration | Internal or external |
Display | LCD |
Operating Temperature | 15°C to 30°C |
Power Supply | 100-240 VAC, 50/60 Hz |
Prerequisites for working on Sartorius high-resolution balances.
Notes on the optimum work area for balance operation and testing.
Information on essential measuring and test equipment for balance checks.
Listing of special tools required for Sartorius Research Balances.
List of relevant documentation for RC/MC balances.
Description of the function keys for MC and RC balance models.
Procedure for setting the balance's identification code, time, and date.
Guidelines for creating GLP/GMP-compliant ID codes.
Instructions for setting the balance's time and date formats.
How to view the balance's hardware and software version numbers.
Accessing and setting the balance's operating menu code.
Conditions required before checking balance metrological specifications.
Guidelines on using test weights and performing balance testing.
Procedure for checking balance reproducibility using standard deviation.
Method for calculating the standard deviation of weighing results.
Procedure to test and verify off-center loading tolerances.
Steps for performing and evaluating internal span adjustment.
Procedure for checking the balance's linearity.
Sequence of adjustments and necessary preparations for balance calibration.
Steps to configure the balance for BPI (Binary Processor Interface) mode.
Procedure for coarse adjustment of off-center load errors.
Procedure for fine adjustment of off-center load errors.
Procedure to verify balance linearity.
Methods for external linearity adjustment of RC/MC balances.
Steps for performing internal linearity adjustment on MC balances.
Specific procedure for MC-balances external linearity adjustment.
Procedure to overwrite internal linearity weights.
Procedure to overwrite internal calibration/adjustment weights.
Activating and performing automatic internal balance adjustment (isoCAL).
Steps to open and securely close the balance housing.
Procedure for exchanging the draft shield assembly.
Instructions for cleaning the balance's weighing system.
Guides for exchanging display, digital, and analog PCBs.
Diagram showing the assembly of balance components.
Diagram illustrating the internal structure and PCBs of the balance.
Guide to identifying and resolving balance error codes and operational issues.
Overview of balance operating menu codes and functions.
Template for recording balance calibration and test results.