EasyManua.ls Logo

Sartorius Micro MC Series - Installing the Draft Shield

Sartorius Micro MC Series
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
RC-F-32.TIF
RC-F-29.TIF
RC-F-32.TIF
RC-F-30.TIF
Installing the Draft Shield
Remove the guide bushing (107) for the new draft shield (111) from the
centering ring (108).
Install the new draft shield (111) and adjust it.
Fasten the new draft shield (111) in this position through the base plate
using the screws (113).
Route the connecting cables (E) of the draft shield drive through the retain-
ers on the enclosure frame (208/211) and insert the plug back into the
socket of the digital PCB (204).
Install the guide bushing (107), align it according to the pan adapter
(430) and tighten it on the centering ring (108).
Important Note:
Use the guide bushing (107) from the old draft shield when installing a new
draft shield on balances of model series RC 250 S up to serial number
10602262. The reason: these types of balances have a thicker pan
adapter (430).
Close the balance housing (see page 24).
Now replace the shield disk (106), shield ring (105), weighing pan
(103) and glass cylinder (102) of the weighing chamber draft shield).
Finally, place the draft shield cover(s) (101 / 104) on the draft shield
(111). Now the balance is ready to operate.