EasyManuals Logo

Sartorius Micro MC Series User Manual

Sartorius Micro MC Series
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
22
RC-A-38.TIF
RC-A-12.TIF
RC-A-37.TIF
RC-A-34.TIF
RC-A-23.TIF
RC-A-34.TIF
Internal Adjustment
Make sure that the balance operating menu is set to code “ 1 10 1
- internal calibration/adjustment accessible -.
Press the key (MC) or the key (RC) for at least 3 seconds when a
stable readout is displayed.
The readout shown on the left will appear.
Press the key, and the internal adjustment weight will be applied by
servomotor and removed at the end of adjustment. The weight value is
stored.
Then the balance returns to the weighing mode (see the Figure on the left).
After the internal adjustment routine the span adjustment has to be tested
externally again.
Automatic Internal Adjustment
Set the balance operating menu to code “1 15 3 “ to activate the self-
calibrating and automatic adjustment function (isoCAL).
When isoCAL is activated, the scale symbol will flash in the display and
the balance will automatically perform internal calibration and adjustment.
The adjustment parameter are: Temperatur changes >1.5K
and every 4 hours
If a printer is connected, a GLP/GMP-compliant record will be printed
out.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Micro MC Series and is the answer not in the manual?

Sartorius Micro MC Series Specifications

General IconGeneral
BrandSartorius
ModelMicro MC Series
CategoryScales
LanguageEnglish

Related product manuals