Internal Adjustment
– Make sure that the balance operating menu is set to code “ 1 10 1 “
- internal calibration/adjustment accessible -.
– Press the key (MC) or the key (RC) for at least 3 seconds when a
stable readout is displayed.
– The readout shown on the left will appear.
– Press the key, and the internal adjustment weight will be applied by
servomotor and removed at the end of adjustment. The weight value is
stored.
– Then the balance returns to the weighing mode (see the Figure on the left).
– After the internal adjustment routine the span adjustment has to be tested
externally again.
Automatic Internal Adjustment
– Set the balance operating menu to code “1 15 3 “ to activate the self-
calibrating and automatic adjustment function (isoCAL).
– When isoCAL is activated, the scale symbol will flash in the display and
the balance will automatically perform internal calibration and adjustment.
– The adjustment parameter are: Temperatur changes >1.5K
and every 4 hours
– If a printer is connected, a GLP/GMP-compliant record will be printed