13
Advanced Functions
Repeated Blow-out, available with Picus® NxT, can be used in conjunction
with Manual Pipetting mode for liquids where normal dispensing does not
completely empty the tip.
For more information on using advanced functions see section 3.2.10.
3.2.5. Diluting
In Dilution mode liquids separated by an air gap are aspirated and then
dispensed simultaneously. Diluting can be used to dilute samples and reagents.
The diluent is aspirated first followed by an air gap, then the sample or
reagent to avoid contamination.
To select Diluting mode:
1. Select MENU by pressing the left Softkey.
2. Use the adjustment wheel to select Diluting and confirm by pressing the
operating button or the right Softkey for OK.
To edit the volume and speed settings:
1. Press the middle Softkey for EDIT or turn the adjustment wheel all the way
to the right or left. The first setting to be edited is now highlighted.
2. Turn the adjustment wheel to set the desired value.
3. Confirm the setting by pressing
a. the operating button or the right Softkey for OK, which will exit the
editing mode.
OR
b. the middle Softkey for NEXT, to move on to edit the next highlighted
setting.
4. Repeat steps 2 and 3 for all settings you wish to edit.
5. To exit the editing mode without saving changes, press the left Softkey for
BACK.
To use the pipette with the selected program:
1. Press the operating button to aspirate the diluent.
2. Press the operating button again to aspirate the air gap.
3. Press the operating button a third time to aspirate the sample.
4. Dispense the entire volume by pressing the operating button.
5. Empty the tip by pressing the operating button again.
6. Eject the tip by pressing the electronic tip ejector button.
Advanced Functions
Repeated Blow-out (Picus® NxT only) and Mixing can be used in conjunction
with Diluting.
−Mixing mixes liquids manually or automatically. The mixing volume can be
adjusted according to the pipette’s maximum volume.
−Repeated Blow-out helps to remove residual liquid from the tip, for liquids
where normal dispensing does not empty it sufficiently.
For more information on using advanced functions see section 3.2.10.
3.2.6. Sequential Dispensing
Sequential Dispensing repeatedly dispenses selected volumes in any desired
order. This is a useful mode for diluting series and making calibration curves.
To select Sequential Dispensing mode:
1. Select MENU by pressing the left Softkey.
2. Use the adjustment wheel to select Seq. Disp. and confirm by pressing the
operating button or the right Softkey for OK. The latest settings used will
then be displayed.