22
Choose the desired target volume
Enter the volume actually measured
To save the adjustment, press either the Softkey for YES or
the operating button.
The difference between
the target volume and the
actual volume exceeds
the limit that can be
calibrated.
The pipette is set to use the
adjustment values saved
to ADJ1.
To change the accuracy adjustment settings:
1. If password protection is enabled, log in using the administrator profile,
(only in Picus® NxT).
2. Select MENU by pressing the left Softkey.
3. Use the adjustment wheel to select Setup and confirm by pressing the
operating button or the Softkey for OK.
4. Highlight Adjustment by turning the adjustment wheel. Confirm the
selection by pressing the operating button or the Softkey for OK.
5. Use the adjustment wheel to select ADJ1, ADJ2, or ADJ3. Press the middle
Softkey to EDIT.
6. Use the adjustment wheel to select 1 Point, 2 Points, or 3 Points. Confirm
your selection by pressing the operating button or the Softkey for OK.
7. For 1-point adjustment, first set the target volume for accuracy
adjustment by turning the adjustment wheel and confirm by pressing the
operating button or the Softkey for OK.
8. For 2 or 3-point adjustment, the target volumes are already set. Just
confirm them by pressing the operating button or the Softkey for OK.
9. After setting or accepting the target volume, insert the actual volume
delivered as per your measurement result by turning the adjustment
wheel. Confirm by pressing the operating button or the Softkey for OK. If
the difference between the target volume and the actual volume exceeds
the limits for calibration, then the display will show “Calibration values
out of range”.
10. Repeat steps 7 and 8 for 2 or 3-point adjustment, until all volumes are set.
11. When the pipette asks if you want to save the adjustment data, confirm
by pressing the operating button or the right Softkey for YES. If you do
not wish to save the data, press the left Softkey for NO.
12. Note that you may exit the editing mode at any time, without saving any
changes by pressing the left Softkey for BACK.
To activate saved adjustment settings:
1. If password protection is enabled, log in using the administrator profile
(only in Picus® NxT).
2. Select MENU by pressing the left Softkey.
3. Use the adjustment wheel to select Setup and confirm by pressing the
operating button or the Softkey for OK.
4. Highlight Adjustment by turning the adjustment wheel. Confirm the
selection by pressing the operating button or the Softkey for OK.
5. Use the adjustment wheel to select ADJ1, ADJ2, or ADJ3. Confirm the
selection by pressing the operating button or the Softkey for OK.
6. When the pipette asks if you wish to set the custom adjustment values,
confirm by pressing either the operating button or the right Softkey for
OK. If you do not wish to set the adjustment, press the left Softkey for NO.
7. If you confirmed the setting, ADJ1, ADJ2, or ADJ3 is now displayed on the
lower left corner of your screen, and the pipette has returned to the Setup
menu. Press the left Softkey for BACK two times to get back to the main
display.