EasyManuals Logo

Sartorius PICUS User Manual

Sartorius PICUS
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
34
Weighing
1. Adjust the desired test volume (VS).
2. Carefully fit the tip onto the tip-cone.
3. Fill the tip with test water and empty it five times to achive humidity
equilibrium in the dead air volume.
4. Replace the tip. Pre-wet the tip by filling it once with the test water and
emptying.
5. Aspirate the test water, immersing the tip only 2-3 mm below the surface
of the water. Keep the pipette vertical.
6. Withdraw the pipette vertically and touch the tip against the side wall of
the test water container.
7. Pipette the water into the weighing vessel, touching the tip against the
inside wall of the vessel just above the liquid surface at an angle of 30°
to 45°. Withdraw the pipette by drawing the tip 8-10 mm along the inner
wall of the weighing vessel.
8. Read the weight in mgs (mi).
9. Repeat the test cycle until 10 measurements have been recorded.
10. Convert the recorded masses (mi) to volumes (vi) by multiplying the mass
with the correction factor Z (Z-values in the table below): Vi = mi Z
Z-values (µl/mg)
Temp. (°C) Air Pressure (kPa)
95 100 101.3 105
20.0
20.5
21.0
21.5
22.0
22.5
23.0
23.5
1.0028
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0036
1.0028
1.0029
1.0031
1.0032
1.0033
1.0034
1.0035
1.0036
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0035
1.0036
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0036
1.0037
11. Calculate the mean volume (V) delivered: V = (Vi)/10
12. For conformity evaluation calculate the systematic error eS of the
measurement:
in µl: eS = V – VS VS = selected test volume
or in %: eS = 100 (V – VS)/VS
13. And calculate the random error of the measurement as standard
deviation:
or as coefficient of variation CV =100s/V
14. Compare the systematic error (inaccuracy) and random error (imprecision)
to the performance specification values of your own laboratory.
NOTE! Systematic error is the difference between the dispensed volume and
the target volume. Random error is the scatter of the dispensed volumes
around the mean of the dispensed volume (ISO 8655-1).
NOTE! Sartorius specifications are achieved in strictly controlled conditions
(ISO 8655-1). Users should establish acceptable maximum permissible errors
based on the field of use and the accuracy requirements placed on the pipette
(ISO 8655-1).

Other manuals for Sartorius PICUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PICUS and is the answer not in the manual?

Sartorius PICUS Specifications

General IconGeneral
TypeElectronic Pipette
Power SourceRechargeable Battery
ChannelsSingle-channel and Multi-channel
ErgonomicsLightweight, ergonomic design
Operating ModesPipetting, diluting
MaterialDurable Plastic and Metal Components
ProgrammableYes
Volume Range0.2 µl - 10 ml (depending on the model)
DisplayColor Screen
AccuracyHigh accuracy and precision, exceeds ISO 8655 requirements

Related product manuals