Setup and Configuration PR 5210 Operating Manual
28 Sartorius
4.8.3 Bus size
The normal bus size is 8.
10 byte bus size is used for a coded data transfer.
Default is 8.
4.8.4 Communication
The serial line could be used for data communication with a remote display (e.g. PR1627 or PR1628) or
a SMA data protocol.
Default is 'off'. See chapter 2.5.3.
4.8.5 Baudrate
The baudrate can be selected between 300, 600 ..... and 19.200, care has to be taken that the same
transfer speed is set at the remote display / terminal. Default is 9600.
See chapter 2.5.3.
4.8.6 Access
The access code can consist of 9 decimal digits at maximum. If the access code is set to 0, no check on
the code is done. If an access code has been set (has to be entered twice for security reason), data and
parameters cannot be altered, without entering the code. As long as the user has got access, he is
allowed to change the access code.
Default is 0.
4.8.7 Outputs
The following selections are possible per each of the 3 digital outputs:
transparent
The PLC is controlling the output
ADU error
Output is set if ADU is in error state
Limit1
Result of comparison of limit 1 values with actual weight (see chapter 4.8.9)
Limit2
Result of comparison of limit 2 values with actual weight (see chapter 4.8.9)
Limit3
Result of comparison of limit 3 values with actual weight (see chapter 4.8.9)
Tare active
Output is set, if transmitter is in net mode
4.8.8 Inputs
The following selections are possible per each of the 3 digital inputs:
none
Input is not used for transmitter control
set zero
The transmitter will be set to zero (see chapter 4.7.2.10).
(standstill has to be fulfilled, else standstilltimeout will be given, see chapter 4.7.2.5 , 4.7.2.7
and 4.7.2.9)
set tare
The transmitter will be switched to net mode.
(standstill has to be fulfilled, else standstilltimeout will be given, see chapter 4.7.2.5 , 4.7.2.7
and 4.7.2.9)
reset tare
The transmitter will be switched to gross mode

Table of Contents