Sartorius PR 5210/10 - 7.4 Analog output; 7.7 Profibus DP; 7.8 Power supply

Sartorius PR 5210/10
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Data PR 5210 Operating Manual
68 Sartorius
7.2 RS 232 interface
RJ 12 connector, functions: Calibration, configuration, operation via WIN-tool on PC.
See chapter 2.5.2
7.3 RS 422 / 485 interface
Terminal block to connect a remote display / PC for SMA protocol.
See chapter 2.5.3.
7.4 Analog output
See chapter Fehler! Verweisquelle konnte nicht gefunden werden.
7.5 Digital inputs
See chapter 2.5.5
7.6 Digital outputs
See chapter 2.5.6
7.7 Profibus DP
Standard EN 50 179 volume 2, PROFIBUS
DIN 19245: PROFIBUS, Process Field Bus (part 1 and 3)
Baudrates 9.6, 19.2, 93.75, 187.5, 500 [kBps], 1.5, 3.0, 6.0, 12.0 [MBps], automatic detection
Buffer size 8 bytes
I/O data 8 bytes
UserPrm No
Sync YES
Freeze YES
Clear YES
Set-Slave-Add No
7.8 Power supply
Power voltage 24 VDC +10% / -15%
Power consumption 8.2 W