Profibus Interface PR 5210 Operating Manual
56 Sartorius
6.3 Profibus Parameter Numbers
With the following parameters all configuration and calibration data can be written to or read from
the transmitter.
All other parameters belong to the class extended configuration, which can be found on separate
pages.
Values which are marked with (factory setting) represent the factory delivery data.
A parameter uses always 4 bytes (format DINT).
6.3.1 Configuration parameter
6.3.1.1 Parameter 1: Analog output
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 Range Output mode
Byte 3 ADU in error state ADU below zero ADU above FSD
Analog output
Description Range / signal factory setting
Output mode off,
(reg. 30)*,
gross,
net
0,
1,
8,
9
0
Analog range 0...20 mA
4...20 mA
0
1
1
ADU in error state** hold
0 mA
4 mA
20 mA
0
1
2
3
1
ADU below zero Linear ***
0 mA
4 mA
20 mA
0
1
2
3
0
ADU above FSD Linear***
0 mA
4 mA
20 mA
0
1
2
3
3
* Output value 1 means that the value contained in register 30 is written to the analog output.
** The ADU error state is only valid if output value is set to gross (8) or net (9).
*** The linear selection can be used to get a proportional output signal outside the range 0 to FSD.
6.3.1.2 Parameter 2: Scaling weight value for 0/4 mA output
The WEIGHT value is stored at which the analog output is giving 0 or 4 mA.
Default = 0
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 WEIGHT MSB
Byte 1 ""
Byte 2 ""
Byte 3 WEIGHT LSB