PR 5210 Operating Manual  Profibus Interface 
Sartorius  57 
6.3.1.3  Parameter 3: Scaling weight value for 20 mA output 
  Bit 7  Bit 6  Bit 5   Bit 4  Bit 3  Bit 2  Bit 1  Bit 0 
Byte 0  WEIGHT MSB 
Byte 1  "" 
Byte 2  "" 
Byte 3  WEIGHT LSB 
 
The weight value is stored at which the analog output is giving 20 mA. 
Default = 3000 
 
 
6.3.1.4  Parameter 4: Communication 
  Bit 7  Bit 6  Bit 5   Bit 4  Bit 3  Bit 2  Bit 1  Bit 0 
Byte 0   
Byte 1   
Byte 2   
Byte 3        COM 2
1
  COM 2
0
 
 
The transmission to a remote terminal (e.g. PR 1628) or the SMA protocol can be switched on or off. 
COM = 0: switched off, 
COM = 1: a remote terminal (e.g. PR 1628) switched on, 
COM = 2: SMA protocol switched on, 
default = 1. 
 
 
6.3.1.5  Parameter 5: Digital outputs 
 
 
Description Value  
transparent* 
aduerr 
Limit1 
Limit2 
Limit3 
tare active 
0 
1 
2 
3 
4 
5 
 
* Transparent mode means, that the PLC is writing to the outputs directly. 
Limit(x) means, that the transmitter is writing the limit result to the outputs directly. 
 
  Bit 7  Bit 6  Bit 5   Bit 4  Bit 3  Bit 2  Bit 1  Bit 0 
Byte 0   
Byte 1      Output 1 , default = 0 
Byte 2      Output 2 , default = 0 
Byte 3      Output 3 , default = 0