PR 5210 Operating Manual Profibus Interface
Sartorius 65
6.3.3.16 Parameter P48: Zerotrack range
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 RANGE MSB
Byte 3 RANGE LSB
RANGE 0 ... 500.00 d default = 0.25 d
Value = d*100
6.3.3.17 Parameter P49: Zerotrack step
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 STEP MSB
Byte 3 STEP LSB
STEP 0 ... 10.00 d default = 0.25 d
Value = d*100
6.3.3.18 Parameter P50: Zerotrack repeat
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 REPEAT
REPEAT 0 ... 100 measuring times default = 0 (Zero tracking is off)
6.3.3.19 Parameter P51: Overload
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 OVL MSB
Byte 1 ""
Byte 2 ""
Byte 3 OVL LSB
OVL Overload in d 0..9999999 default = 9 d