EasyManuals Logo

Sartorius Tacta User Manual

Sartorius Tacta
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #4 background imageLoading...
Page #4 background image
4
SafetySpace
air gap
]
1.3 Sartorius Pipette Tips
We recommend only using Sartorius pipette tips with the Tacta® pipette. Using fully
compatible tips guarantees the pipette’s performance specifications and ensures pipetting
accuracy and precision.
Sartorius tips are made of pure virgin polypropylene and manufactured in protected
cleanroom conditions.
Sartorius non-filter tips are available in trays, bulk packages, and space-saving refill systems.
Both non-filter tips and trays are autoclavable at 121°C (252°F), 20 min, 1 bar (15 psi). All
Single Tray racks and clean Refill Packs are certified RNase, DNase, and endotoxin-free.
To avoid aerosol contamination, use SafetySpace Filter Tips and discard them after each
pipetting. These can also be used to avoid sample loss caused by the sample coming into
contact with the filter. The extra space between the sample and the filter (see picture)
ensures that even foaming or viscous liquids do not come into contact with the filter in
reverse pipetting or repetitive/multiple dispensing modes.
For more information, please visit www.sartorius.com or contact your local Sartorius
representative.
2. Package Contents
The Tacta® pipette package contains the following items:
Tacta® mechanical pipette
Sartorius Optifit tip(s)
Adjustment tool
Four identification tags
Safe-Cone Filters (except 3 and 10 μl models)
Autoclavable grease, 1 g
Pipette holder
Carousel stand adapter
QC certificate from gravimetric testing according to ISO 8655-6 standard
Quick Start Guide
If any item in the package is missing or damaged, contact your local Sartorius
representative.
1.4 Adjustment Tool
For situations where factory calibration is not applicable, the adjustment tool allows you to
calibrate the pipette accurately.
NOTE: For more information on adjusting of the Tacta® pipette see 8.2, Adjusting the pipette.

Other manuals for Sartorius Tacta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Tacta and is the answer not in the manual?

Sartorius Tacta Specifications

General IconGeneral
BrandSartorius
ModelTacta
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals