4
SafetySpace
air gap
]
1.3 Sartorius Pipette Tips
We recommend only using Sartorius pipette tips with the Tacta® pipette. Using fully
compatible tips guarantees the pipette’s performance specifications and ensures pipetting
accuracy and precision.
Sartorius tips are made of pure virgin polypropylene and manufactured in protected
cleanroom conditions.
Sartorius non-filter tips are available in trays, bulk packages, and space-saving refill systems.
Both non-filter tips and trays are autoclavable at 121°C (252°F), 20 min, 1 bar (15 psi). All
Single Tray racks and clean Refill Packs are certified RNase, DNase, and endotoxin-free.
To avoid aerosol contamination, use SafetySpace Filter Tips and discard them after each
pipetting. These can also be used to avoid sample loss caused by the sample coming into
contact with the filter. The extra space between the sample and the filter (see picture)
ensures that even foaming or viscous liquids do not come into contact with the filter in
reverse pipetting or repetitive/multiple dispensing modes.
For more information, please visit www.sartorius.com or contact your local Sartorius
representative.
2. Package Contents
The Tacta® pipette package contains the following items:
− Tacta® mechanical pipette
− Sartorius Optifit tip(s)
− Adjustment tool
− Four identification tags
− Safe-Cone Filters (except 3 and 10 μl models)
− Autoclavable grease, 1 g
− Pipette holder
− Carousel stand adapter
− QC certificate from gravimetric testing according to ISO 8655-6 standard
− Quick Start Guide
If any item in the package is missing or damaged, contact your local Sartorius
representative.
1.4 Adjustment Tool
For situations where factory calibration is not applicable, the adjustment tool allows you to
calibrate the pipette accurately.
NOTE: For more information on adjusting of the Tacta® pipette see 8.2, Adjusting the pipette.