SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Chapter 1
Operating Elements
Automatic Mode (ECG) Settings (cont.)
Interpretation (C version only)
To print or suppress interpretation statements on the printout proceed as follows:
Interpretation (Interpretation Option Only)
Entry Key Sequence Printout Confirm
ALT
1 or
2
4
5
Interpretation is
omitted
Press
STOP
key
6
Interpretation is
printed
Confirm the selection by pressing the STOP key
STOP
Full details of the interpretation option are given in the SCHILLER ECG Measurement and
Interpretation booklet (Art. No. 2.510 179).
Interpretation Settings (C version only)
The interpretation settings enable the user to determine whether or not certain comments will be
added to the interpretation statements on the ECG printout. Furthermore, the patient’s age can be
defined (<30 or >30) and if low or high sensitivity should be applied. Low sensitivity will suppress
certain nonspecific ECG diagnosis; this may be advisable when carrying out ECGs for screening.
Interpretation Settings
Entry Key
Sequence
Setting Confirm
ALT 6
1
"Normal" / "Abnormal" is not
printed
Press
STOP
key
2 "Normal" / "Abnormal" is printed
3 "Unconfirmed report" is not printed
4 "Unconfirmed report" is printed
5 Patient age assummed to be < 30
6 Patient age assummed to be > 30
7 Low sensitivity
8 High sensitivity
Note: The `Patient age assumed to be..` setting is only applicable when patient data has
not been entered.