2 General Safety Information
2.2 Product Safety
2.2.4 Combination with Medical Electrical Equipment
The device can be combined with components from other manufacturers provided
that all components are compliant with the medical electrical equipment safety re-
quirements according to IEC 60601-1.
It is the operator’s responsibility to check and make sure that the system is and
remains fully operational.
When using devices from different manufacturers and when operating an endo-
scope and/or endoscopic accessories with electromedical devices, it must be en-
sured that the applied part is properly insulated (in accordance with the require-
ments for type BF/CF applied parts defined in IEC 60601-1).
Contact our Service Department if you have any questions regarding compatibility.