7 Product Data
7.1 Technical Data
7 Product Data
7.1 Technical Data
Dimensions (W x H x D) 295 x 130 x 355 mm
Weight 8 kg
Current consumption 1.6 - 0.6A
Supply voltage 100-240V~, 50/60Hz
Protection class acc. to IEC 60601-1 Protection class I
Device fuse T1.6AH 250V
Interfaces 2x MIS-bus
Degree of protection IP 21
Applied part as per IEC 60601-1 Type CF defibrillation-proof
Classification as per IEC/CISPR 11 Group 1 Class B
7.2 Ambient Conditions
Transport and storage conditions
Temperature -20 °C to +70 °C
Relative air humidity 5 % to 95 %
Atmospheric pressure 50 kPa to 106 kPa
Operating conditions
Temperature +10 °C to +40 °C
Relative air humidity 10 % to 90 %
Atmospheric pressure 70 kPa to 106 kPa
CAUTION
Non-observance of the ambient conditions
Irreversible tissue damage to patient or undesirable coagulation, user injury,
material damage
Observe operating conditions as well as transport and storage conditions
en
Instructions For Use | 13 - Mar - 2020 | Version: B Page 73 of 78