4 Operation
4.3 Technical Inspection Prior to Use
4.3 Technical Inspection Prior to Use
4.3.1 Visual Inspection
Performing a visual inspection
A visual inspection must be performed prior to each medical procedure.
Proceed as follows:
Is the housing
undamaged?
1. Inspect the housing of the light source for external damage.
Do not use the light source if there is any external damage to the housing.
Are the electric cables
undamaged?
2. Check that all electric cables are undamaged and positioned safely.
Do not use the device if the electric cables are damaged or present a trip or
entanglement hazard.
Is the light guide
undamaged?
3. Check that the light guide you intend to use is undamaged.
Do not use the light guide if it is broken or twisted.
This concludes the visual inspection.
4.3.2 Function Check
Performing a function check
A function check must be performed prior to each medical procedure.
Proceed as follows:
1. Connect the light guide.
2. The light guide should be directed away from the eyes to avoid dazzling
anybody.
3. Switch on the light source.
Check that you are able to adjust the light intensity and that the system is
functioning correctly.
Do not use the light source if you are unable to adjust the light intensity.
This concludes the function check.
en
Instructions For Use | 13 - Mar - 2020 | Version: B Page 63 of 78