What to do if... ProLab 2000
100
ba75577e03 11/2009
Warning [TempErr]
Warning [ISEErr]
6.3 Conductivity measurement
Conductivity measuring
cell was not recognized
Error message, Error
(invalid
calibration)
CalClock flashes
6.4 D.O. measurement
D.O. sensor was not
recognized
Cause Remedy
Temperature difference
between measurement and
calibration greater than 2 K.
Adjust the temperature of
the test sample
Cause Remedy
Electrode voltage outside
calibrated range
Recalibrate
Cause Remedy
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and replace it if
necessary
Unsuitable control standard Use control standard,
0.01 mol/l KCl
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Sensor not connected Connect probe
Cable broken Replace probe