EasyManua.ls Logo

SCHOTT Instruments ProLab 2000 - Options for Data Transmission

SCHOTT Instruments ProLab 2000
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProLab 2000 Operation
91
ba75577e03 11/2009
4.10.3 Options for data transmission
Via the RS 232 interface, you can transmit data to a PC or external
printer. The following table shows which data is transmitted to the
interface in which way:
Note
The following rule applies: Except for the menus, the display content is
generally downloaded to the interface with <PRINT> (displayed
measured values, measurement datasets, calibration records).
Data Control Operation / description
Current
measured values
of the connected
sensor
Manual z With <PRINT>.
z Simultaneously with every
manual storage process (see
page 81).
automatic, at
intervals
z With <PRT
_>. Then you
can set the transmission
interval.
z Simultaneously with every
automatic storage process
(see page 82).
Stored measured
values
Manual z Displayed dataset:
with <PRINT> after recall
from data storage.
z All datasets according to the
filter criteria via the Output to
RS232/USB function.
For details, see page 84.
Calibration
records
Manual z Calibration record of a
sensor:
Calibration record of a sensor
with <PRINT> (after calling
up from the data storage or at
the end of a calibration
procedure).
z All calibration records after
calling up from the data
storage via the Output to
RS232/USB function.
For details, see page 87.
automatic z For the respective sensor at
the end of a calibration
procedure.

Table of Contents