EasyManua.ls Logo

SCHOTT Instruments ProLab 2000 - Page 57

SCHOTT Instruments ProLab 2000
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProLab 2000 Operation
57
ba75577e03 11/2009
Preparatory activities Perform the following preparatory activities when you want to calibrate:
Carrying out an
ISE calibration
Proceed as follows to calibrate the meter:
Display Calibration
record
Magnitude of the slope [mV]
+++ 50.0 ... 70.0 or 25.0 ... 35.0
Error Error
< 50 or > 70
or
< 25 or > 35
Eliminate the error according to
chapter 6 W
HAT TO DO IF...
(page 97)
1 Connect the ISE combination electrode to the meter.
The pH/mV/ISE measuring window is displayed.
2 Keep the standard solutions ready.
3 Measure the temperature of the standard solutions using a
thermometer.
1 In the measured value display, select the ISE measuring
window with <><> and <MODE>.
2 Start the calibration with <CAL>.
The calibration display appears.
3 Thoroughly rinse the electrode with distilled water.
4 Immerse the electrode in standard solution 1.
5 Using <><>, select the Set standard setting and press
<MENU/OK> .
6 Set the concentration of the standard solution with < ><>
and press <MENU/OK>.
i Immerse sensor in std. 1
Set temperature: 24 °C
Continue
Set standard: 0.010 mg/l
ISE <CAL>

Table of Contents