ProLab 2000 Operation
69
ba75577e03 11/2009
4 Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5 Press <MENU/OK>.
The measurement with stability control to determine the cell
constant starts.
6 Wait for the end of the measurement with Stability control or
accept the calibration value with <MENU/OK>.
The determined cell constant is displayed as a message (
i).
i Cell constant = 0.475 1/cm
i Immerse sensor in solution
Continue
<CAL>
i Cell constant = 0.481 1/cm
i Temperature =24.8 °C
Terminate stability control
<CAL>
i Cell constant = 0.481 1/cm
i Sensor+++
Accept
<CAL>