ProLab 2000 Operation
93
ba75577e03 11/2009
Cond The following settings for conductivity measurements are reset to the
default settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
respective measuring menu. To open it, activate the relevant
measuring window in the measured value display and press the
<MENU/OK> key.
Oxi The following settings for D.O. measurements are reset to the default
settings with the Reset function:
Setting Default settings
Cal. interval 150 d
Measured parameter
Cell constant (c) 0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of
the linear temperature
compensation
2.000 %/K
TDS factor 1.00
Setting Default settings
Cal. interval 14d
Comparison meas. Off
Measured parameter D.O. concentration
Relative slope (S
Rel
)1.00
Salinity (value) 0.0
Salinity (function) Off