EasyManua.ls Logo

SCHOTT Instruments ProLab 2000 - Page 98

SCHOTT Instruments ProLab 2000
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
What to do if... ProLab 2000
98
ba75577e03 11/2009
No stable measured
value
Obviously incorrect
measured values
Buffer solutions:
Incorrect buffer solutions
Change buffer solutions
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change buffer solutions
Cause Remedy
Cause Remedy
pH electrode:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low Use a suitable electrode
Temperature too high Use a suitable electrode
Organic liquids Use a suitable electrode
Cause Remedy
pH electrode:
Not connected Connect electrode
Cable broken Replace electrode
pH electrode unsuitable Use a suitable electrode
Temperature difference
between buffer and test
sample too high
Adjust temperature of buffer
or sample solutions

Table of Contents