Vers. 20210510ENG Page 34
environments.
Voltage dip, short
interruption and voltage
change on the power
input line
GB/T17626.11
0%Ut, Continue 1
cycle
(On Ut, 100% sag)
0%Ut, Continue 1
cycle
(On Ut, 100% sag)
Network power supply
should have the quality
of use in typical
commercial or hospital
environment.
If the users of this
product need to run
continuously during
power interruption, it is
recommended to use
un-interruptible power
supply or battery power
supply.
40%Ut, Continue 5
cycles
(On Ut, 60% sag)
40%Ut, Continue 5
cycles
(On Ut, 60% sag)
Network power supply
should have the quality
of use in typical
commercial or hospital
environment.
If the users of this
product need to run
continuously during
power interruption, it is
recommended to use
un-interruptible power
supply or battery power
supply.
70%Ut, Continue 25
cycles
(On Ut, 30% sag)
70%Ut, Continue 25
cycles
(On Ut, 30% sag)
5%Ut, Continue 5s
(On Ut, 95% sag)
5%Ut, Continue 5s
(On Ut, 95% sag)
EMF
(50/60Hz) GB/T 17626.8
3A/m
3A/m
If abnormal work
occurs, it is necessary
to keep the product
away from the power
frequency magnetic
field or install a
magnetic shield in the
place. The power
frequency magnetic
field in the anticipated
installation site should
be measured to meet
the requirements below
the level.
NOTE: Ut refers to the AC network voltage before the voltage is applied.