Disabling the safety messages is
not recommended.
DigiPROBE ERROR MESSAGES
Error messages pertaining to the DigiPROBE are
displayed on the run screen in the Temp fi eld. If the
fi eld displays “ERROR” during a run, the system will
stop heating. Power down the controller, check the
connections, turn the power back on. (Remember
that if you are not using the DigiPROBE, the
DigiPROBE dongle must be connected.) If problem
persists, consult the service department (see page
85).
Disconnecting or reconnecting the
umbilical cable while the system is
turned on, especially while the run
screen is displayed, will likely result
in permanent damage to the
instrument.
DIAGNOSTICS TEST PROCEDURE
If you are having trouble with the DigiPREP
controller, follow the steps below and contact the
service department.
1. Shut off power to DigiPREP keypad.
2.
Ensure that DigiPROBE is installed and
inserted into either the holding well or the
sample. If you are not using DigiPROBE,
ensure that the dongle is connected.
3. Ensure that the umbilical cable is connected
properly to the block and the controller.
4. Turn the power back on.
5. Is the blue LED light on? YES NO
6. Is the LCD backlight lit up? YES NO
7. If there is a problem with the LCD, describe
the appearance of the LCD after booting:
FAINT SQUARES BLANK
Please phone your SCP service department
(see page 85) and report this information
before returning the product. This will help our
technicians determine the cause of the malfunction
more quickly.
CLEANING
The keypad may be cleaned using a mild soap,
water, and a clean cloth. Ensure that the controller
is disconnected from the power supply. No organic
solvents should be used as it may destroy the fi nish
on the keypad mask. The system should not be
immersed in any liquid.