KeyPad
Version 1.0
SCP SCIENCE
39
OPERATION WITH PROBES
DigiPROBE
DigiPROBE (optional) may
be used with DigiPREP
system to monitor and control
the sample temperature (when
placed in the sample) or the
block temperature (when
placed in the storage well
located on the block).
When using the DigiPROBE with the
DigiPREP HP, a fl at-bottomed, conical
fl ask is recommended for best results.
Install the PROBE in the port marked 'DigiPROBE'
on the back of the DigiPREP block. If using to
monitor and control the heating of the block via the
sample temperature, insert the PROBE end into a
control sample with the PROBE holder supporting
the DigiPROBE. (The 010-500-036 PROBE holder
is used with 50ml DigiTUBEs and 010-501-036
is used with 100ml DigiTUBEs and a 6"(15 cm)
PROBE. The 010-500-037 PROBE holder is used
with glass digestion tubes and a 10" (25 cm) PROBE.)
If using to monitor & control the block
temperature, place the DigiPROBE in the storage
well, located at the rear right corner of the graphite
block. Ensure that there is a glass liner (catalogue
number 010-500-041) in the well.
The control sample should be a typical
digestion sample. The tip of the
DigiPROBE should be approximately
1 mm from the bottom of the digestion
vessel.
Serious temperature overrun may
occur during digestion if the DigiPROBE
is not in an appropriate sample or
the storage well. Always place the
DigiPROBE in the storage well when
not in use.
DigiPROBE can be calibrated with DigiPREP
controllers or the PC Classic controller (with
Go2Temp software). Please see your controller
manual for calibration instructions.
DigiSET
DigiSET (optional) can be used with all DigiPREP
low temperature systems to monitor and suspend
the heating cycle at a pre-set sample volume level.
DigiSET PROBE is connected to the DigiSET unit
and 'COM 2' from DigiSET unit is connected
to the 'COM 2' of the controller with the supplied
cable. Please consult the DigiPREP controller manual
for more detailed instructions.
NOTE
NOTE
CAUTION