40 •
4.2 Set up scales
The scales are fully assembled on delivery.
ATTENTION!
Incorrect measurement with force shunt
If the housing of the scales is resting on some-
thing, e.g. a towel, the weight measurement will
not be correct.
When setting up the scales, ensure that they
only make contact with the ground via the
foot screws.
1. Place the scales on a flat, stable surface.
2. Align by turning the foot screws.
5. OPERATION
5.1 Weighing
WARNING!
Serious injury to baby following a fall
Baby scales generally stand on raised work sur-
faces. If the baby falls from this surface, this
may result in serious irreversible or fatal injuries.
Never leave a baby unsupervised on the
scales.
Start weighing 1. Check that there is no load on the scales.
2. Press the Start key.
The display will read se[A and then quickly run
through all elements of the display.
The scales are ready for operation when the display
reads
0.000.
If the device is being operated with a power supply
unit, the symbol will appear in the display.
3. Place the baby on the scales.
4. Read off the measured result.