9. MAINTENANCE
9.1 Information about maintenance
Your seca scales leave the factory with an accuracy
greater than ± 0.3%. To preserve this level of accuracy
the product must be set up carefully and serviced regu-
larly. We recommend having it serviced every 3 to 5
years depending on how often the scales are used.
ATTENTION!
Incorrect measurement due to improper
maintenance
Make sure that maintenance and repair are
only carried out by an authorised service
partner.
You can find your local service partner at
www.seca.com or just send an e-mail to
service@seca.com.
Er:X:12 is displayed?
The scales have been switched on with too great
a load.
- Reduce load on scales
- Restart scales
Er:X:16 is displayed?
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Do not touch the trolley or table supporting
the scales
- Restart scales
… when the Enter key
(send/print) is pressed
and
Er:X:71 is dis-
played?
Data transmission not possible, wireless module is
disabled.
- Activate wireless module (see "Activate
wireless module (SYS)" on page 49).
… when the Enter key
(send/print) is pressed
and
Er:X:72 is dis-
played?
Data transmission not possible, no wireless group
set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 50).