54 •
bAtt is displayed?
Batteries are flat.
- Insert new batteries
St0p is displayed?
The maximum load has been exceeded.
- Reduce load
is displayed?
The ambient temperature of the scales is too high
or too low.
- Set up scales in an ambient temperature
between +10 °C and +40 °C
- Wait approx. 15 minutes until scales have
adjusted to ambient temperature
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
The device was unable to send measured
results to the radio receiver (seca wireless prin-
ter or PC with seca USB wireless module).
Check that the scales are integrated in
the wireless network.
Check that the receiver is switched on.
Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.
... only the item "SYS" is
visible in the rf menu?
The wireless module is disabled.
- Activate wireless module (see "Activate
wireless module (SYS)" on page 49).
... only the items "SYS"
and "lrn" are visible in
the rf menu?
The wireless module is activated but no wireless
group set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 50).
... the items "APrt" and
"Time" are not visible in
the rf menu?
No wireless printer is logged into the wireless
group.
- Use menu item "lrn" to log the wireless printer
into the wireless group(see "Set up wireless
group (Lrn)" on page 50).
Er:X:11 is displayed?
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight
more evenly
- Restart scales
Malfunction Cause/Remedy