Cleaning • 53
English
2. Select the menu item "Time" in the "rf" submenu.
3. Confirm the selection.
The current selection for the year (Year) is dis-
played.
4. Select the correct year.
5. Confirm the selection.
6. Repeat steps 3. and 4. as appropriate for the
month ( ), day (
dAy), hour (hour) and minute
().
7. Confirm your selection each time.
After confirming your selection for Minute you will
leave the menu automatically.
These selections will be automatically transmitted
to the wireless printer.
The wireless printer automatically adds the date
and time to every printout.
NOTE:
For further operation of the wireless printer see
its user manual.
7. CLEANING
To clean the tray and housing of the scales use a
household detergent or standard disinfectant as
required. Take note of the instructions provided by the
manufacturer.
8. WHAT TO DO IF …?
Malfunction Cause/Remedy
… no weight is dis-
played during weighing?
The scales are not connected to the power supply
- Check the scales are switched on
- Check batteries are inserted
0.000 does not appear
before weighing?
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
… a segment lights up
continuously or not at
all?
There is a fault at that point.
- Notify maintenance service.
is displayed?
Battery voltage is falling off.
- Insert new batteries as soon as possible