INSTRUCTION MANUAL CODE 80342 REV O 11/2021 Page: 56
J.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
Check drain valve.
Check and clean the outlet lter.
Check the connection of the drain hose.
23 POWERFAILPowercutoffwiththeequipmentoperating
Cause: During an operation cycle, the power has failed.
Resolution: Restart the cycle.
24 POWERFAILRESUMABLE
Cause: It indicates a cut off in the power while the equipment was operating.
The process can continue because the device has a battery and has preserved the status.
(If the temperature conditions during the cut have varied too much, another error may
appear, such as 36, as the chamber temperature has dropped too low).
29 SAFETYTHERM Safetyperformanceoftheboilerthermostat
Cause: Inner temperature has reached a too high value.
Resolution: • Wait for a few minutes and reset the thermostat
• Increase parameter "B" in 5s.
30 LEVELERROR Waterlevelisnotdetected
Cause: No water enters the chamber.
Resolution: • Check the water inlet tap
• Increase parameter "B" so that more water enters.
34 INTERNALUNDERTEMP Readingwaterprobetemperature(connectorplate:
K20)lowerthan8ºCformorethan5minutesduringtheHEATphaseorinany
momentduringanyotherphasewhentheprogramisoperating.
Cause: • If this happens at the beginning of the cycle, it is likely that the water in the network is
very cold (winter).
Resolution In the autoclave loaded with cans, pour half a bucket of hot water into the boiler and
then start the autoclave.
36 STERILIZ.UNDERTEMP Temperaturedrop
Cause: In the sterilization phase, the temperature has dropped 2ºC below the set point for 30s.
Resolution: Check the temperature probe and the location inside the jar.
Make sure there has been no power interruptions.
37 STERILIZ.OVERTEMPExcessivetemperature
Cause: In the sterilization phase, the temperature has exceeded the set point at 4ºC for 5s.
Resolution: Only for 20L. Increase parameter D.
38 EMPTYPRESSURETimeout
Cause: The temperature has been above the cooling temperature and the pressure has been
below the cooling pressure for 60 minutes.
39 STERILIZATIONPRESSURE
Cause: During the sterilization phase we have not reach the purge pressure in 30 seconds.
Resolution: Check that the back tap is not open, that the equipment is not leaking, that no solenoid
valve is open.
40 DISCONNECTED
Cause: Communication with the power plate has been lost.
Resolution: Turn off the equipment and check connection between circuits.
41 PREHEATtimeoutExcessivewaterpre-heatingtime
Cause: The preheat temperature is not reached before 60'.
Resolution • Check (Reset) the safety thermostat.
• Check the heating elements.
42 PURGEHEATtimeout Thepurgetemperatureisnotreachedbefore60'
Resolution: • Check (Reset) the safety thermostat.
• The product is very cold, increase the temperature of the PRE phase.