5
Attach the male nut (C) of the pump
outlet tube to the line filter (A) by
hand.
6
Fix the line filter (A) with a 10 mm
wrench (B) and tighten the male nut
(C) of the pump outlet tube with an 8
mm wrench (D).
When using in the front end of
LC-MS, carry out the cleaning
process described as follows to
reduce background noise. Pump
50% methanol solution at a
pressure of approx.10 MPa for
approx. 30 min. (For example,
attach a stainless steel tube
with ID 0.1 x 2000 mm (Part
No.: 228-57908-41) and adjust
the flow rate.)
116
LC-40D X3 CL LC-40D XR CL
4 Maintenance