www.Siargo.com MF5806E1 User Manual 25 | Page
Reset timer
0x00F2
Write
Y
Read
N
Description
Reset the meter timer
Value type
UINT 16, Fixed value 0x0003
Notes
To reset the timer of the accumulated therapy time/duration.
Note: need to disable the write protection.
Write protection
0x00FF
Write
Y
Read
N
Description
Write protection disabler for a set value to a specific register.
Value type
UINT 16, Fixed value 0xAA55
Notes
This function is enabled at the time of product shipment. To enable the write
function
of a specific parameter, such as offset calibration, reset accumulated
flow rate
, or reset the timer, the user needs to send 0xAA55 to the register
0x00FF, and then the write function will be enabled (write protection is
disabled). After the write execution is completed, the firmware will
automatically re-enable the write protection.
Note: Only Address and Baud rate will not be write-protected.