Name Description
[MF1 Distance] Hyste‐
resis [mm]
Enter the distance threshold hysteresis see "MF1 – Dist submenu",
page 38.
Options
•
1 ... 300,000 [mm]
Factory setting
•
10 [mm]
[MF1 Speed] Threshold
[mm/s]
Enter speed threshold. The switching hysteresis is fixed at 100 mm/s.
Options
•
0 ... 15,000 [mm/s]
Factory setting
•
5,000 [mm/s]
[MF1 Speed] Mode Select direction for detecting the speed overrun.
Options
•
negative direction [-]
•
positive direction [+]
•
both directions [+/-]
Factory setting
•
both directions [+/-]
[MF1 Service] Laser
Warning
The output switches when the aging of the laser diode means that the
device is nearly due to be replaced.
Options
•
off
•
on
Factory setting
•
on
[MF1 Service] Level
Warning
The output switches when the signal attenuation value reaches the criti‐
cal value, for example due to contamination.
Options
•
off
•
on
Factory setting
•
on
[MF1 Service] Temp.
Warning
The output switches when the interior device temperature reaches the
critical value.
Options
•
off
•
on
Factory setting
•
on
[MF1 Service] Plausib.
Warning
The output switches if extrapolated measured values are to be output for
implausible measured values see table 21, page 44, ErrRej.
This may be caused by an internal error, hardware errors, obstruction of
the light beam, optical interference or electrical interference.
Options
•
off
•
on
Factory setting
•
on